Model BTDS03 Water Quality Tester Calibration Instructions

Membrane Solutions® handheld meters are factory-calibrated and designed to stay consistent. If you do need to recalibrate your meter follow the instructions below:

Items Required:

  • Membrane Solutions® BTDS03 Digital TDS EC Meter.
  • Standard Calibration Solution 500ppm.
  • Clean Cup/Container.

Calibration Method

  1. Rinse the cup/container with a small amount of the calibration solution and empty to drain
  2. Refill the cup with calibration solution
  3. Hold TDS meter in the solution up to the maximum immersion level, lightly stirring meter to dislodge any pockets of air from the sensor.
  4. Press the ON/OFF button once to turn on the meter, wait for a further 15-seconds for the reading to stabilize in the solution.
  5. Press and hold the MODE button for 3 seconds until the TDS reading on the screen starts to flash, then release the button.
  6. After 2-3 seconds, the electronic calibration to 500ppm (± 2%) will complete, and the TDS reading will stop flashing.

Membrane Solutions Water Quality Tester Instructions  & troubleshooting 2The calibration is now complete.
Membrane Solutions Corp
3320 W Valley HWY North Ste D108
Auburn WA 98001

How to Make your own TDS Calibration Solution

If you prefer to make your own 500ppm Standard TDS solution suitable for calibrating the Membrane Solutions® BTDS03 Digital TDS EC Meter, please follow these  instructions:

Items Required:

  1.  1 Litre of pure water (H2O). RO or distilled water is OK.
  2. 0.25g of table Salt (NaCI).
  3.  Mixing container, >1 Litre.

Method

  •  Carefully measure out 0.25g (250mg) of plain table salt.
  • Mix the table salt into the liter of water until it’s completely dissolved, careful not to spill any water as it must still be 1
    Litre in volume for it to maintain the correct strength.

Calibration Tip: When calibrating the Membrane Solutions TDS meter, try and maintain the temperature of the calibration solution as near to 25 o C (77°F) as possible, this is the temperature reference point and will give a more accurate calibration result.

FAQ


Setting up Membrane Solutions Water Quality Tester Instructions is easy with these step-by-step instructions. Follow the steps are you are set-up in no time.

Using Membrane Solutions Water Quality Tester Instructions is simple by using this manual. Follow the instructions one by one and everything will be fine

Is Membrane Solutions Water Quality Tester Instructions stuck or not working properly? By resetting it you can set it back to the standard settings and everything should be back to normal. This works in many cases when problems arise.

This page contains the complete user instructions & manual for Membrane Solutions Water Quality Tester Instructions

All symbols, colors and codes for Membrane Solutions Water Quality Tester Instructions are explained in this instruction manual. You can find everything you need and moree